You are Here:  Tech Center > Expert Advice > ICP Multi-Element Solution Incompatibilities
  General Applications
 
Certified Concentration of a Standard
Nitric Acid Oxidations
Two Methods for Certification
ICP-MS Sensitivity Optimization
Cleaning Quartzware
 
  Sample Preparation
 
Dissolving Graphite to Measure by ICP
Quartz Powder Preparation
XRF vs. ICP for Concrete Metals
Dissolution of Si in Metals
Analysis of Sediment Samples
Reactions of Sulfur and Bromine
Determination of As in Sediment
Matrices for Soil Preparation
Trace Metal Analysis of Air Condensate
Analyzing Cement
Determination of Hg in Sediment
Sample Size for Steel Analysis
Analyzing For Br by ICP
Analytical Tips for Tin
Analyzing Silicone Oils
Ashing Technique for Metals
Coal Sample Digestions
Ashing Procedure For Tin
Using Blue HDPE Containers
Sn and Tl Soil Extraction
Filter Papers for EPA Method 353.2
Determination of Hg in Biological Matrices
 
  Stability & Compatibility
 
Eliminating High and Erratic Blanks
HF in Multi-element Standards
Chemical Stability of Standards
Questions About Mo, Ti, and Hg
Overcoming Au Stability Issues
Contaminants of Ethanol in HDPE
Vegetable Oil Stability
ICP Multi-Element Solution Incompatibilities
Multi-Element Compatibility and Preparation
Stability of Mercury in Plastic Bottles
Cadmium, Lead, and Copper Stability
Diluted Blends of Single Element Standards
Incompatible Metals Within Solution
Stability of Oil Standards
 
  Measurement
 
Controlling Transpiration
Calculating Weighing Uncertainty
Expressing Sensitivity
As Line Interferences from Fe
Measurement Uncertainty
Positive Interference with Hg 202
Linearity Slope Variance
Mercury and Antimony Troubleshooting
Zn Interferences Upon the Na 330.237 nm Line
Aluminum, Beryllium, and Chromium Interferences
 
ICP Multi-Element Solution Incompatibilities Technical Troubleshooter

Who is Paul Gaines?

Karen's department was having some difficulty with their ICP-OES getting a full list of elements to pass at 1 mg/L for their criteria of ± 10% recovery. When they prepared their lab mix at 1 ppm using a stock standard of 23 elements (Ag, Al As, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, K, Mg, Mn, Mo, Na, Ni, Pb, Sb, Se, Tl, V, and Zn all at 100 ug/mL), the addition of other elements such as P, S, Si, Sn, Li, and Ti yielded many inaccurate readings. Karen wondered if there might be interference.


IV Labs / Inorganic Ventures: White Background
 
From the desk of Paul Gaines, Ph.D. Dr. Paul Gaines
  RE: ICP Multi-Element Solution Incompatibilities
Dear Karen,

     When you add the P, S, Si, Sn, Li, and Ti, there is a lot of chemistry that may take place. The list is as follows:

1) The S will react with the Ba and Pb causing a precipitate, assuming that the manufacturer uses sulfate. IV uses a different sulfur compound that does not precipitate with any of these elements.

2) Sn and Ti can partially hydrolyze out of solution depending upon how these elements are chemically "fixed" (i.e. - are they fluorides or chlorides).

3) If Cl is present, Ag will precipitate or photoreduce depending upon the Cl and Ag concentrations.

4) If too much Fluoride is present then Ba and Ca can partially precipitate.

5) The viscosity of the solution may be altered sufficiently to cause a bias in the readings. The use of internal standardization is often used to compensate for this type of interference.

We have designed solutions and identified spectral lines that will give accurate readings for all of the elements in the periodic table. The only potential problem could be that the instrument you're using MAY not have the spectral range or sensitivity.

Let me know if I can be of further help.

 
  Paul's signature
Paul R. Gaines, Ph.D.
Serving you in chemistry
 
IV Labs / Inorganic Ventures: White Background

An Inorganic Ventures feature - December 2002


DISCLAIMER  Advice offered by Dr. Gaines is intended for the individual posing the question. We urge you to contact us to verify whether these suggestions apply to your circumstances.